Page 17
ITM-MAGMETER-T500-ED2
G1. Datalogger setting without time mask
Use cursor keys to select the option DATALOGGER from main menu and confi rm your selection by EXE
1.
key. A submenu will be displayed.
Select option MASK and confi rm your selection by EXE key.
2.
A time mask is displayed. The time mask is disabled if it is in format xx - xx - xx xx : xx (day – month
3.
– year hrs : min). If any position shows character different from x, select command EDIT and use cursor
keys to correct it to x. Use command RETURN to return to submenu.
Select option VAR and confi rm your selection by EXE key. Select in menu one of optional values to be
4.
recorded (immediate fl ow rate Q, maximum fl ow rate during chosen interval MAX, minimum fl ow rate du-
ring chosen interval MIN or average fl ow rate during chosen interval AVERAGE). Selected option will be
marked with asterisk *. Use command RETURN to return to submenu.
Select option PER and confi rm your selection by EXE key. Use command EDIT to enable interval editing
5.
in minutes. Enter interval length and use command RETURN to return to submenu.
Select START. Option RUN should be marked with asterisk *; this means that datalogger is operating.
6.
Select option RETURN to fi nish setting.
7.
The instrument will ask if you want to save the new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save the new parameters and switch to normal display mode. If NO is selected, parameters will
not be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.
Sample recording can be set with fi xed interval from 15 to 107 minutes (e.g. every 15 minutes after datalogger
start) regardless to real time or using a time mask (e.g. datalogger starts each Friday at 15:00 and measurement
interval takes 2 hours).