38
BASIC OPERATIONS
BA
SI
C OPERATIONS
PREPARATION BEFORE M
EASUREM
E
NT
7
Place the patient's chin on the chinrest and check that his/her forehead is touching to the fore-
head rest.
8
Press the
button to adjust the chinrest height until the eye height mark of the
chinrest reaches the same height as the patient's eye.
SETTING THE SAFETY STOPPER
1
Select the right/left eye by tapping the
button/
button.
CAUTION
• Before measuring, set the safety stop to prevent the measuring
window glass from hitting the patient's eye.
Set it respectively for the right and left eyes.
• Set the safety stop from the side of the instrument.
Setting operations from other positions, it is not easy to check
positions of the eye and ocular pressure measuring window, may
cause injury by touching ocular pressure measuring window to a
patient.
UP/DOWN
Patient ID
Patient ID
/ 3
/ 3
/ 3
/ 3
1
1
TOPCON
TOPCON
T/P
Please touch
pupil on the screen.
Up/Down button for chinrest
Eye height mark
R
L
Patient ID
Patient ID
/ 3
/ 3
/ 3
/ 3
/ 3
/ 3
/ 3
/ 3
1
1
TOPCON̲TAROU
TOPCON̲TAROU
MODE
T/P
Please touch
pupil on the screen.