CAUTIONS
Ensuring the Safety of Patients and Operators
To avoid injury when operating the up/down button for the chinrest, be careful not to catch the patient's fingers.
The light emitted from this instrument involves potential risk; the longer the irradiation time, the more risk of dam-
age to the eye.
When the instrument operates with the maximum light volume, exposure for more than 2 hours will exceed the
safety guideline.
If a symptom of photosensitive epilepsy appears in a patient during photography, please stop photography imme-
diately.
Preventing Electric Shocks and Fires
To avoid injury by electric shock, do not open the cover. For repair, call your service engineer.
Electromagnetic Compatibility (EMC)
This instrument has been tested (with 100/120/230V) and found to comply with IEC60601-1-2:Ed.3.0:2007. This
instrument radiates radio frequency energy within standard and may affect other devices in the vicinity. If you
have discovered that turning on/off the instrument affects other devices, we recommend you change its position,
keep a proper distance from other devices, or plug it into a different outlet. Please consult your authorized dealer
if you have any additional questions.
Содержание SP-1P
Страница 1: ...USER MANUAL SPECULAR MICROSCOPE SP 1P ...
Страница 2: ......
Страница 18: ...16 COMPONENTS 5 Tap the to return to the capture screen check that patient ID is updated Next button 1081 ...
Страница 25: ...23 COMPONENTS SETUP SCREEN ...
Страница 90: ......
Страница 92: ...SPECULAR MICROSCOPE SP 1P 41904 97030 Printed in Japan 1308 100LW0 ...