9-5
9 CHECK AND ADJUSTMENT
9.3.1 Checking /Adjusting the Plate Level
Adjustment is required if the axis of the plate level is not perpendicular to the vertical axis.
l
Check
1) Place the plate level parallel to a line running through the centers of two leveling screws, say, A and
B. Use these two leveling screws only and place the bubble in the center of the plate level.
2) Rotate the instrument 180° or 200g around the vertical axis and check bubble movement of the
plate level. If the bubble has been displaced, then proceed with the following adjustment.
l
Adjustment
1) Adjust the level adjustment capstan screw, with the accessory adjusting pin and return the bubble
towards the center of the plate level. Correct only one-half of the displacement by this method.
2) Correct the remaining amount of the bubble displacement with the leveling screws.
3) Rotate the instrument 180° or 200g around the vertical axis once more and check bubble
movement. If the bubble is still displaced, then repeat the adjustment.
9.3.2 Checking /Adjusting the Circular Level
Adjustment is required if the axis of the circular level is also not perpendicular to the vertical axis.
l
Check
1) Carefully level the instrument with the plate level only. If the bubble of the circular level is centered
properly, adjustment is not required. Otherwise, proceed with the following adjustment.
l
Adjustment
1) Shift the bubble to the center of the circular level, by adjusting three capstan adjustment screws on
the bottom surface of the circular level, with the accessory adjusting pin.
Leveling screw A
Plate level
Leveling screw B
Level adjustment
capstan screw
Half amount of displacement
Capstan adjustment
screws
Bottom of the base
Содержание GTS-821A, GTS-822A, GTS-823A, GTS-825A
Страница 1: ...INSTRUCTION MANUAL AUTO TRACKING TOTAL STATION GTS 820A SERIES GTS 821A GTS 822A GTS 823A GTS 825A...
Страница 2: ......
Страница 120: ...11 4 11 SETTING ATMOSPHERIC CORRECTION...
Страница 121: ...11 5 11 SETTING ATMOSPHERIC CORRECTION...
Страница 143: ......