TROUBLESHOOTING
The error messages are displayed with a red or yellow frame on the touch display.
When the error message is displayed, please tell the message ID to service engineers of your dealer
or TOPCON listed on the back cover.
The message ID is displayed in [ ].
Messages during operation
Error message
Contents
GALVANO error.
Please turn the power switch OFF.
[010107]
It is probable that the galvano unit does not operate prop-
erly.
Turn off the power and contact the service engineer.
Overcharged Xenon flash.
Please turn the power switch OFF.
[111102]
Overcharge has occurred in the xenon lamp (fundus photog-
raphy light source) power supply.
Turn off the power and contact the service engineer.
Internal temperature error.
Please turn the power switch OFF.
[101111/109111/110111]
The temperature in the instrument is abnormally high.
Turn off the power and contact the service engineer.
Light volume exceeds upper limit.
Please turn the power switch OFF.
[100112]
The laser light intensity is beyond the upper limit value.
Turn off the power and contact the service engineer.
Failed to measure internal temperature.
Please turn the power switch OFF.
[109114]
An error in the communication with the temperature sensor
has occurred.
Turn off the power and contact the service engineer.
Failed to measure light volume.
Please turn the power switch OFF.
[100115]
Measuring the laser light intensity has failed.
Turn off the power and contact the service engineer.
Light volume is lower than lower limit.
Please turn the power switch OFF.
[100116]
The laser light intensity is below the lower limit value.
Turn off the power and contact the service engineer.
Internal communication error.
Please turn the power switch OFF.
[209001/210001/211001/212001]
An error in the communication between the built-in PCBs
has occurred.
Turn off the power and contact the service engineer.
Failed to initialize the built-in image board.
Please turn the power switch OFF.
[201002]
Initialization of the image board has failed.
Turn off the power and contact the service engineer.
Camera is not connected.
Color fundus photography is disabled.
[202001]
The connection with the color photography camera has
failed. Turn off the power and contact the service engineer.
Camera is not connected.
FA/Auto Fluo photography is disabled.
[203001]
The connection with the FA/FAF photography camera has
failed. Turn off the power and contact the service engineer.
Failed to initialize a camera.
Color fundus photography is disabled.
[202002]
An initialization error of the color photography camera has
occurred.
Turn off the power and contact the service engineer.
Failed to initialize a camera.
FA/Auto Fluo photography is disabled.
[203002]
An initialization error of the FA/FAF photography camera
has occurred.
Turn off the power and contact the service engineer.
Failed to charge Xenon flash.
Please turn the power switch OFF.
[211101]
An error has occurred while the xenon lamp (fundus photog-
raphy light source) power supply is being charged.
Turn off the power and contact the service engineer.