37
BASIC OPERATIONS
BA
SIC OPE
R
ATIONS
FU
NDUS TOMOGRAPH
Y
Setting up the patient
1
Peel off one chinrest paper. If there is not chinrest paper, supply it. (Refer to P.91.)
2
Wipe off dust on the forehead rest.
3
Seat the patient comfortably on an exam stool or chair in front of the instrument.
4
Adjust the table height or chair height so the patient can relax with his/her chin placed cen-
trally on the chinrest.
Let the patient rest his/her chin on the chinrest.
5
Adjust the chinrest height by adjusting the
button so the outside corner of the
patient's eye is level with the Canthus marker on the chinrest post.
Let the patient rest his/her forehead on the forehead rest.
CAUTION
• To avoid electric shock, do not touch the external connection terminal,
the cable terminal from the external device and the patient at the same
time.
• When moving the chinrest up and down, be careful not to pinch the
patient's hand to avoid possible injury.
• When moving the main unit, the external fixation target will be pushed
by the measuring head. To avoid injury of the patient, be careful to
prevent the external fixation target from hitting the patient.
NOTE
• If the patient wears glasses or contact lenses, have him/her remove them first.
CHINREST UP/DOWN
Canthus marker