3-8 Replacement of the PCB for the near-point illumination LED
Operation status
Operation procedure
1. Remove the near-point rod and near-point chart.
2. Remove the hidden screws and four screws of the
cover at the patient side, using the Phillips
screwdriver. Then, remove the cover.
3. Remove the two screws from inside.
4. Remove the parts at the inspector side. Remove
the hexagonal screw of the forehead rest knob from
the side, using a hexagonal wrench (for 6SU3).
5. Remove the connector. Remove the old PCB,
using the precision Phillips screwdriver.