52
−
Photo of operation
Operation procedure
Writing
15. Select the ROM file with “OPEN”.
16. Press [START].
If writing starts though the specified ROM file is not
the same as the ROM file of the instrument specified
in “Target Model”, an error dialog box appears. In
this case, check the setting and then open the file
again.
17. Writing starts. (The near-point illumination LED of
the instrument lights up.)
Checking
18. When writing is finished, the result of sum check is
printed through the compact power supply.
19. Make sure that the result of sum check is the same as
the value of the technical instruction manual.
20. Turn off the power of the compact power supply.
21. Set DIPSW of the compact power supply to the
original condition. Close the cover of the compact
power supply.
22. Set the cover on the instrument.
R-CV5000-0-0704-01