17. CHECK AND ADJUSTMENT
17-13
2) Manual Adjustment Procedure
Operating procedure
Operating key
Display
Press the [MENU] key from normal
measurement mode.
[MENU]
MENU
1/1
F1 : PARAMETERS
F2 : PROGRAMS
F3 : REMOTE
Select [F1:PARAMETERS].
[F1]
PARAMETERS
1/1
F1 : TRK
F2 : MEAS
F3 : COM
Select [F1:TRK] .
[F1]
TRK
1/3
F1 : SEARCH
F2 : WAIT TIME
F3 : SCAN WIDTH
Press the [F4] key, TRK menu 2/3 will
be shown.
[F4]
TRK
2/3
F1 : ADJ AXIS(M)
F2 : ADJ AXIS(A)
F3 : TRK SPEED
Select [F1:ADJ AXIS(M)].
[F1]
ADJ
AXIS
(M)
PRISM
COLLIMATE
SET
OK
[YES]
– – –
Adjust H/V jog shuttle until collimating
proper prism center.
Press the [F3] (YES) key.
[F3]
ADJ
AXIS
(M)
AUTO
TRACKING
SET
OK
[YES]
– – –
Press [A/M] key, tracking the prism.
[A/M]
ADJ
AXIS
(M)
AUTO
TRACKING
SET
OK
[YES]
– – –
Confirm that tracking the prism, and
press the [F3] (YES) key.
1)
The measured data will be checked by
AP-L1A.
[F3]
ADJ
AXIS
(M)
AUTO
TRACKING
DATA CHECK