SPECIFICATIONS
GENERAL
Load Battery Test Current.............................................
About 300mA
Battery Voltage Display Accuracy..................................
±
0.5V
BATTERIES
Type alkaline ................................................................
1.5V
Other Type....................................................................
1.2V
Quantity.........................................................................
8
DISPLAY
Type..............................................................................
LCD
2 lines x 16 Characters
AUTO-OFF
Automatic Turn OFF Time after last action...................
5Min.
VOLTMETER (3 minutes maximum)
Automatic Voltage DC Range.......................................
0 to 950Vdc
Automatic Voltage AC Range........................................
0 to 700Vac
Resolution on both voltage types..................................
±
1V
Accuracy on both voltage types....................................
±
1.5% of FS
INSULATION TEST
Resistance Range
250V Test Voltage.........................................................
2k-2G Ohms
500V Test Voltage.........................................................
4k-4G Ohms
1000V Test Voltage.......................................................
8k-8G Ohms
Accuracy.......................................................................
>1M
W
:
±
3%rdg
30K
W
-1M
W
:
±
5%rdg
<30K
W
:
±
20%rdg
Typical Output Voltage @ 1mA
250V Setting ................................................................
275V Max.
500V Setting ................................................................
550V Max.
1000V Settings ............................................................
1100V Max.
When Voltage is constant, Current is software
Limited at.......................................................................
±
1.2mA
Short Circuit Current on all ranges.
Maximum Short Circuit Current.....................................
±
4mA.
Polarization Index
Ratio Resolution............................................................
±
0.1
Ratio Accuracy..............................................................
±
1% of Rdg