DIGITAL TORQUE GAUGE MODEL
ATGE-G
8
Procedures of Various Settings
Measurement display
This chaper explains various settings.
* Before setting, make sure that the counter shows 000 (RUN
mode).
Press MD for voer 2 seconds till the mode setting
display appears. Then, release the MD key.
8-1. Unit setting
Selected torque unit
(cN·m
mN·m
gf·cm
ozf·in
cN·m)
Press to save the setting and proceed to the next.
Press to cancel. The display retur ns to the
measurement display.
Proceed without saving.
Unit conversion table (Conversion factors)
8-2. PEAK mode change
PEAK mode change
P-1: PEAK measuring mode
Captures the peak and holds the value.
If you continue to apply torque, measured value will be
renewed.
P-2: Torque screwdriver measuring mode.
Holds the torque value upon clicking of the torque
screwdriver.
Press to save the setting and proceed to the next.
Press to cancel. The display returns to the measuring
display.
Proceed without saving.
mN·m
10.000
gf·cm
101.972
ozf·in
1.4162
(1) Unit setting
Unit
"P-2" selected
(2) PEAK mode change