DIGITAL TORQUE GAUGE MODEL
ATGE-G
· Holding the chuck holder fixed, open the three-jaw chuck and put the torque screwdriver bit in it.
· Place the torque screwdriver bit in the center of three-jaw chuck. While holding the chuck
holder fixed, tighten the chuck, and ensure that the bit is clamped completely.
· Set the counter to 001-999 for peak hold (P-2).
· While holding the torque gauge fixed, turn the torque screwdriver and measure the torque.
· When measuring a rotary slip type screwdriver, measure one click (within 1 sec).
Fixing Plate (Option plate)
Chuck holder
Display
Fix the torque gauge.
Three-jaw check
Make sure to hold the chuck holder when rotating the three-jaw chuck.
Turn the torque screwdriver.
Note * Make sure to hold the chuck holder when rotating the three-jaw check, otherwise the
torque may transfer to the main body directly and damage the product.
* Make sure to fix the measuring object in the center and keep it level against the torque
gauge in order to measure the accurate torque.
* In case that auto-zero be effected when the measuring object is set to ATGE-G, make
sure the object and ATGE-G is in stable condition.
* Hold the screw driver upright when measuring.
* Avoid applying any load on the chuck holder.
* The torque setting of the screwdriver must not exceed ATGE-G maximum capacity.