2.3
General Specifications
2.3.1
Performance
Active Degrees of Freedom (system of 3+ isolators)...................................................................................6
Active Bandwidth...................................................................................................................... 0.6 to 150 Hz
Transmissibility at Resonance ............................................................................................................... <1.1
Isolation above 2Hz.............................................................................................................................. >90%
Settling Time
2
(10 lb. Step input) .....................................................................................................<300 ms
Dynamic Range........................................................................................................................... 60 dB Max.
Static Load Capacity/Isolator ........................................................................... 182-2045 kg (400-4500 lbs.)
Number of Isolators / system ......................................................................................................... 3 or more
Maximum Displacement................................................................................15
µ
m peak-peak below 10 Hz
Maximum Static Load (non-operating)............................................................................2727 kg (6000 lbs.)
2.3.2
Physical Dimensions
DC-2000 Controller ...............................................482.6mm(19”)W x 368.3mm(14.5
″
)D x 44.5mm(1.75”)H
..........................................................(1 Standard RETMA unit w/rack mounting holes)
Isolator Size ..........................................300mm(11.75”)W x 320mm(12.5”)D x 277mm(10.9”)H (unloaded)
2.3.3
System Requirements
2.3.4
Environmental
(refer to EN 61010-1: 1993, EN 61010-1/A2: 1995):
For indoor use only, up to an elevation of........................................................................... 2,000m (6560ft.)
Maximum allowable operating temperature range:.............................................. 10°C to 32°C (50 to 90°F)
Maximum allowable storage temperature range:.............................................-40°C to 55°C (-40 to 130°F)
Maximum allowable humidity .................................................................. 53% up to 40°C, (20°C dew point)
Tolerance in mains supply voltage.......................................................................+/-10% of nominal voltage
Over voltage category ..................................................................................................................................2
Pollution degree (IEC 664)...........................................................................................................................2
Controller ventilation requirements .................................. 25mm clearance on sides, 0mm top and bottom
2.3.5
Floor Requirements
Flatness....................................................................................... +/- 1.5 mm from perfect plane at isolators
Stiffness (low load isolators) .............................................................................4.4x10
7
N/m (250,000 lb./in)
Stiffness (medium load isolators) .......................................................................6.2x10
7
N/m (350,000 lb./in)
Stiffness (high load isolators) ...............................................................................8x10
7
N/m (450,000 lb./in)
2.3.6
Utility Requirements
Mains Power Supply ............................................................... 90 – 230VAC, 50 or 60 Hz, less than 400VA
Computer (optional) ........................................................... PC compatible w/ a 1:1, M-F, DB-9 serial cable
2
Settling time is defined as the time required for the oscillation amplitude to decrease by 90% from the initial peak amplitude