STACIS
2100 Piezoelectric Isolation System - Installation & Operation Manual
P/N 96-28765-01
Page 33 of 43
7. Turn the controller power OFF, and then back ON after 10 seconds. The System Status
LED on the Controller’s front panel will be “
red”
, and then turn “
green”
and start to blink.
Click the
OK
button above or press the
Enter
key on the computer keyboard immediately
after the status LED starts blinking
“green”
.
The LED on the Controller’s front panel will turn “
amber”
and start blinking rapidly,
indicating that the download is in progress. The
Serial Loader
window will now indicate the
progress of the download in the
Status
window.
8. After the download is complete, the
Status
box will indicate the total bytes transmitted, and
the time required in seconds. ‘Successful’ means the downloaded .HEX file passes the
checksum test, and is written to flash (non-volatile) memory.
After the download is complete, the LED will turn “
green”
, and the controller will automatically
reboot, and the new software will start to run.
•
New software download resets all system parameters to the factory default settings.
Therefore, the system must be re-configured for the specific application/installation.
Parameters such as number of isolators, control gains, etc. typically require
adjustments; refer to the
Description of Menu Tree Functions
section for instructions.
•
After configuring the system, permanently save the parameters in memory using the
“Save Parameters” menu option.
Download
Progress
Download
Result