Methods
Measurement
Fixthe adapter for 16 mm vials on the sample chamber.
Switch the unit on using the [ON/OFF]key.
The display shows the following:
Select the required test using the [MODE] key.
ScrollMemory (SM)
To avoid unnecessaryscrolling for the required test range, the instru-
ment memorizesthe last range used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test range first.
The display shows the following:
Placethe
blank
in the adapter (Note 1–4) making sure that the
marks
are aligned.
Blanks are specially prepared for each individual test range.
Press the [ZERO/TEST]key.
The "Range" symbol flashes for approx. 8 seconds.
RANGE
RANGE
On
Off
Zero
Test
Mode
COD vario
Select the appropriate vial for the desired range:
LR-Range:
0 –
150 mg/l, Order code 2420720
MR-Range:
0 –
1500 mg/l, Order code 2420721
HR-Range:
0 – 15000 mg/l, Order code 2420722
Preparing the Sample
Open a reaction vial with a white cap and add the specified volume.
(Ensurethat appropriate personnel safety equipment is used.)
LR-/MR:
2 ml water sample
HR:
0.2 ml water sample
Prepare a
blank
(Note 1) by using deionised water (TOC-free) instead of the sample
(LR/MR:2 ml, HR: 0.2 ml).
Replacethe cap
tightly
. Invert the vial gently several times to mix the contents
(The vial
will become hot during mixing!)
and digest the vials for
120 minutes
in the reactor at
a temperature of
150°C
. Removethe vials from the reactor and allow them to cool down
to 60°C or less. Mix the contents by inverting each vial severaltimes while still warm. Then
allow the vials to cool to ambient temperature before measuring.