32
drops in the midway, quickly click the interface pause
command (the stirring sleeve is at the lowest position at
this moment), a. Check whether the stirring sleeve is at the
central position of the deep-hole-plate square hole. If not,
adjust the parameters of motor No. 3 in accordance with
Section 6.2 of the maintenance manual. b. Check whether
there is gap between the stirring sleeve and the bottom of
deep-hole plate. If it is not at the specified position, adjust
the parameters of motor No. 1 in accordance with Section
6.2 of the maintenance manual. c. Run the magnetic
absorption command and click the pause command when
the magnetic bar drops in the midway (the magnetic bar is
at the lowest position). Ensure that the magnetic bar is at
the bottom of the stirring sleeve and is not closely attached
to the bottom of the stirring sleeve. If it is not at the
specified position, adjust the parameters of motor 35 in
accordance with the maintenance manual.