axial section of the motor. Install it on the rotating shaft of the 42 motor at the direction as shown in the
figure, and then place it on the fixed height-limiting tooling of the synchronous belt pulley according to
the method as shown in the figure. Tighten two jackscrews to fix the belt pulley P5 (the jackscrews need
to be coated with 243 medium-strength thread glue. After jackscrews are tightened, keep them still for
more than 8 hours until they are solidified), as shown in the figure below.
4.
If steps 2 or 3 have been carried out, the alignment position of the instrument must be checked as
follows:
1) Take at least 4 deep-hole plates, place them at positions 1, 3, 4, 6 of the experiment module and correctly
place the stirring sleeve at columns 2 and 8;
2) Enter the debugging mode, select the horizontal pushing motor column, click the test button at the hole 2
position, and run the heating module to the corresponding position;
3) Select the lifting motor column, click the test button at the top of the stirring sleeve, and run to the
corresponding position;
4) Check whether the stirring sleeve is aligned to the rotating sleeve from both sides of the instrument. If
not, slightly adjust the position parameters of the horizontal pushing motor hole 2 until the stirring
sleeve is aligned to the rotating sleeve as shown in the figure below;
5)
Make a confirmation, click “Save” and “Backup Parameter” (only for new version), restart the
instrument and follow up the experiment results.
Horizontal pushing
motor
Horizontal pushing motor
synchronous belt pulley
Jackscrew
Height-limiting
tooling of
synchronous pulley