42
man_vd6s2-de-130515
set
5s
Examples:
Action
Telegram to VD6
Answer telegram
from VD6
Read pressure measurement
"001M^
C
R
"
"001M864022K
C
R
"
(
8.64x10+2mbar
= 864 mbar)
Start Control
"001a1c
C
R
"
"001a1c
C
R
"
Read setpoint 2
"001S2V
C
R
"
"001S400018Q
C
R
"
(
4.0x10-2 mbar)
Unlock setpoint 2
Set setpoint 2 to
4.2x10-2 mbar
"001s2v
C
R
"
"001s420018s
C
R
"
"001s2v
C
R
"
"001s420018s
C
R
"
Unlock gas-correction-factor 1
Set gas-correction-factor 1
to 1.20
"001c1e
C
R
"
"001c000120W
C
R
"
"001c1e
C
R
"
"001c000120W
C
R
"
Before writing the parameters "s" (setpoint), "c" (correction factor) and "h"
(hysteresis) they must be activated by sending the related unlock command!
For further information please refer to the separate description of the Thyracont
communication protocol.
5 Configuration
Changes of the instrument configuration should be carried
out only by qualified personnel!
In order to change settings of the VD6 in the configuration mode:
Hold the "set"-key pressed while the
VD6 is connected to mains supply
until the display shows "typ ..".