3.4 Transducer Connection 0-10 V (VD12)
The instrument must be switched off before any transducers are
connected. The use of original accessory cables is recommended.
Disregarding this instruction may lead to damage of the instru-
ment.
To this port Analogline transducers for absolute pressure with 0 - 10 V signal out-
put can be connected. The VD12 automatically detects the transducer type and
provides voltage supply for the transducers.
Amphenol
Pin 1:
Pin 2:
Pin 3:
Pin 4:
Pin 5:
Pin 6:
C91E, 6-pole,female, bolted-type
Identification
n.c.
Signal input, 0-10 V
GND
24 VDC, max. 8 W
AGND
Transducers connected to the analog signal inputs have a fixed assignment to
channel 1 or channel 2.
If an analog transducer connection is used, the corresponding
channel(s) cannot be applied as an address for digital trans-
ducers at the RS485 port!
Examples:
channel 1:
analog
channel 2 can be assigned to RS485
channel 2:
analog
channel 1 can be assigned to RS485
channel 1, 2: analog
no channels for RS485 available
1
2
3
4
5
6