LFS4100
Chapter 3: Description
Rev B, July 24, 2018
Page 5
3.5.1. External Workstation Connections
The splicer workstation is the main component of the LFS4100 system. The PC and gas supply, as well as
peripheral components such as an optical power meter, are connected to its back panel as illustrated below.
Figure 3
Splicing Workstation Connections
Power:
The LFS4100 workstation is powered from an external dual-voltage supply. Connection is made via a 5-pin round
connector.
Communication:
PC – LFS4100 Control communication:
RS-232 9-way D-Sub connector, Connect to Com 1 on the PC
Camera - PC:
IEEE-1394 6-pole connection, Connect to IEEE1394 4 pole on
PC
Note:
The computer and the LFS4100 have to be powered off before connecting and disconnecting the camera
firewire cable between the LFS4100 and the computer.
Analog Outputs:
2 BNC connectors - Analog 1 and Analog 2
Gas:
Argon:
1/8” push-fit connector
Vacuum:
5/32” push-fit connector