Press "Save" to save the user calibration. SPLICCO will notify you in case the
new calibration would lead to negative or other non valid wavelengths.
Note
: SPLICCO uses a polynomical fitting routine to create the pixel-wavelength
correlation array. Make sure that the calibration points are distributed over the
whole pixel range.
Figure 53
Wavelength calibration window
3.19
Persistence
SPLICCO offers the possibility to change the persistence attributes for each
panel. You can reach the menu either by right clicking on a window and choosing
"Persistence..." from the appearing menu or selecting 'View / Persistence' from
the main menu.
The persistence dialog offers two sliders for duration and intensity of the
persistence feature. After you have chosen the paramters, press "Apply" to make
the changes valid. Press "Done" to leave the dialog.
Содержание SP USB Series
Страница 1: ...2008 Operation Manual Thorlabs SPx USB USB 2 0 Fiber Optical Spectrometer ...
Страница 2: ...Version Date 1 0 08 Feb 2008 2008 Thorlabs 2008 Thorlabs ...
Страница 6: ...General Information Splicco Part I ...
Страница 9: ...Installation Splicco Part II ...
Страница 20: ...Operating Instruction Splicco Part III ...
Страница 46: ...Virtual devices Splicco Part IV ...
Страница 50: ...Service and maintenance Splicco Part V ...
Страница 53: ...Specifications Splicco Part VI ...
Страница 55: ...53 Specifications 2008 Thorlabs 6 1 Dimensions Figure 60 SPx USB dimensions ...
Страница 56: ...Certifications and compliances Splicco Part VII ...
Страница 58: ...Appendix Splicco Part VIII ...
Страница 65: ......