Ø1/2" Shutters and Controller
Chapter 4: Setup and Operation
Page 8
TTN020243-D02
Chapter 4
Setup and Operation
4.1. Connection Guide
1. Remove components from the case.
2. Secure the SHB05(T) shutter head to your setup using the three
mounting holes, the SM05 thread, or a mounting adapter (sold
separately).
3. Connect the AC power cord to the power supply and then connect to the
AC power outlet. The SHB05 and SHB05T are each supplied with an
external AC-DC switching power supply rated at 100 – 240 VAC, 47 –
63 Hz, and 5 VDC @ 2.4 A maximum output.
4. Place the controller on a flat surface. Connect the mini-DIN connector
from the shutter head, with the arrow on the connector pointing up, into
the controller.
5. Turn the unit on by pressing the power button.
4.2. Manual Shutter Operation
There are two LEDs on top of the unit. The LED next to the power button is green
when the power is on or off when the power is off. The LED next to the enable
button indicates the status of the controller as shown in the table below.
Status LED Color
Meaning
Off
Shutter is Closed
Green
Shutter is Open
Amber
Controller is in Unprotected Mode
and Shutter is Open
Red
There is an Error
Note that when the shutter is closed and in unprotected mode, the LED will be
off. Additionally, unplugging the shutter from the wall outlet will not change its
mode, so if it is in Unprotected Mode and unplugged, it will still be in Unprotected
Mode when plugged back in again. See Section 4.3.2 for more information on
Unprotected Mode.
The shutter generally defaults to the closed position. The following table outlines
the various conditions that will change the shutter's state. For more information
on the meaning of the LED colors and modes, please see Section 4.6.2.