4.8. Typical Performance
The time it takes to open the shutter or close the shutter (IA and DE as shown in
Figure 5) are constant characteristics of the device. The time delay for both
opening and closing is a result of the time it takes to energize the coils.
The net effect is the delay from the initiation of the starting pulse to the shutter
responding (IA) is typically 8 ms. Similarly, the closing time (DE) is typically 7 ms.
The mechanical rise time of the shutter is given as AC in Figure 5. The fall time is
given as EG. The typical values for the rise and fall times respectively are 8 ms
and 9 ms.
Therefore, the total time it typically takes for the shutter to open would be IC or
16 ms. The total time it typically takes for the shutter to close would be DG or
16 ms. Depending on your application, some or all of these values are important
to note.
4.9. SHB05T Coating Performance
The plot above gives the typical reflectance of the PTFE coating on the blades of
the SHB05T. This data is not guaranteed and should be used only as a
reference.