Ø1/2" Shutters and Controller
Chapter 7: Cleaning and Maintenance
Rev F, December 9, 2018
Page 17
Chapter 7
Cleaning and Maintenance
The SHB05(T) shutter and controller can be cleaned using a soft cloth. If needed,
the cloth can be dampened with some isopropyl alcohol. Avoid using any
solvents on or near the unit as well as water as this may cause the blades to rust.
The SHB05(T) incorporates five shutter blades. These blades are thin (0.003")
and lightweight making them very delicate. Please do not insert any objects into
the shutter where the blades may be hit. This may cause a blade to be slightly
bent hampering performance.
Do not place the shutter head close to any magnetic and ferrous materials. It
may affect the performance of the coils within the shutter. Maintain a distance of
at least 1/8" from any ferrous materials. While both sides are susceptible, the
front face is more susceptible to interference.