TIME CLOSED (OFF) 

 

When the active window indicator is in 

TIME CLOSED

 quadrant (lower left) the 

 

 

keys will increment 

and decrement the indicated digit. The 

◄►

 keys will move between digits. This indicates how long the 

shutter will be closed (when the shutter is not connected it indicates the “low” time of the TTL output 
trigger). 

 

NOTE

: If active window indicator is in the 

TIME CLOSED

 quadrant when the shutter is enabled, timing will begin 

with the programmed 

TIME CLOSED

 

 
 

INTERLOCK and ALARM

 

 

Using the Safety Interlock - Connections and Operation 

 

The SC10 is equipped with a Remote Interlock connector located on the rear panel (Refer to Figure 1).  In order 
to enable the controller, a short circuit must be applied across the terminals of the Remote Interlock connector.  In 
practice this connection is made available to allow the user to connect a remote actuated switch to the connector 
(i.e. an open door indicator). The switch (which must be normally open) has to be closed in order for the unit to be 
enabled. Once the switch is in an open state and an SH05 is connected, the shutter will automatically close. And 
the display will flash “

INTERLOCK OPEN

”. If the switch returns to a closed condition the shutter must be re-

enabled at the unit by pressing the 

ENABLE

 switch. 

 

All units shipped from Thorlabs are configured with a shorting device installed in the Interlock connector.  If you 
are not going to use this feature then you can leave the shorting device installed and the unit will operate normally 
as described in the procedures above. 

 

If you wish to make use of the Interlock feature you will need to acquire the appropriate connector mate and wire 
it your remote interlock switch.  Next, remove the shorting device by unscrewing it from the input and install the 
connector into the Interlock input. 

 

The Interlock input only accepts a 

2.5mm mono phono jack

.  This connector is readily available at most 

electronics stores (Radio Shack, Digikey, Mouser, Allied to name a few). 

 
 
 
The electrical specifications for the Interlock input are as follows: 
 
Type of Mating Connector: 2.5mm mono phono jack 
Open Circuit Voltage: +5VDC with respect to Chassis ground 
Short Circuit Current: 0.5mADC 
Connector Polarity: Tip is +5V, Barrel is ground 
Interlock Switch Requirements: Must be N.O. dry contacts (under no circumstances should any external 
voltages be applied to the Interlock input) 

 

8654-D02  Rev. B  8/15/05 

Page 7 of 12 

 

Figure 2

 

- Remote Interlock Connector

 

ALARM 

The SC10 has an Alarm feature when used in conjunction with an SH05 shutter. In the event that the shutter does 
not close within 30 ms of the deactivation voltage (Shutter Closed State) the display will indicate an “ALARM” 
condition. This feature is only active when the SH05 shutter is connected. 

 

Содержание SC10

Страница 1: ...ntroller Operating Manual POWER SHUTTER CONTROLLER SC10 ENABLE O ENABLED TIME OPEN SEC OPEN SHUTTER MODE TIME CLOSED SEC TRIGGER SELECT THORLABS Inc Ph 973 579 7227 435 Route 206N Fax 973 383 8406 Newton NJ 07860 USA ...

Страница 2: ...ctrical and Electronic Equipment Directive 12 The Thorlabs SC10 is a digital shutter controller designed to control the Thorlabs SH05 Shutter The SC10 also has the ability to operate as a stand alone digital delay generator with 1 millisecond resolution and 0 1 millisecond accuracy Features Automated computer control capability 0001 second accuracy On Off times from 1 millisecond to 999 999 millis...

Страница 3: ...riggering other devices or for monitoring the performance of the shutter When the controller is used without a shutter the TTL output signal represents the precise programmed output of the controller A key lock and an interlock function are also provided on the SC10 The SC10 provides drive output pulses that have 0 001Second resolution and 0 0001 Second accuracy for both the ON OPEN and OFF CLOSE ...

Страница 4: ...10 7R 6S Interlock jack 2 5mm mono phono AC Input jack IEC320 Size Length Width Height Weight 11 5 292mm 5 3 135mm 3 0 76mm 5 lbs S N RS232C INTERLOCK 100 240VAC 50 60HZ POWER 11 VA MAX TRIGGER OUT TRIGGER IN TTL TTL MANUFACTURED BY NEWTON NJ USA 435 ROUTE 206 SH05 SC10 REAR PANEL 8654 D02 Rev B 8 15 05 Page 4 of 12 0 0 0 5 N I A U O T T TIME CLOSED SEC TIME OPEN SEC SHUTTER CONTROLLER SC10 O POWE...

Страница 5: ...LED Illuminates when the unit is active or waiting for a trigger event to executing the programmed parameters SHUTTER Illuminates when the shutter is open or the output pulse is active high OPEN OPERATION Initial Set up Locate the unit on a dry level working surface Make sure the POWER switch on the front of the unit is in the OFF position 0 Plug the female end of the AC line cord provided into th...

Страница 6: ...executed the number of times shown next to the REP display The unit may be programmed to repeat the cycle anywhere between 1 and 99 times While the active display is in the REP mode the number of events may be modified by using the key to place the curser under the number displayed and using the keys to increment or decrement the number The repeat sequence begins with activation of a trigger TRIGG...

Страница 7: ...its shipped from Thorlabs are configured with a shorting device installed in the Interlock connector If you are not going to use this feature then you can leave the shorting device installed and the unit will operate normally as described in the procedures above If you wish to make use of the Interlock feature you will need to acquire the appropriate connector mate and wire it your remote interloc...

Страница 8: ...10 can be controlled using the SC10 Application Software provided with the unit It can also be controlled directly using a command line interface and any Terminal Communications software such as HyperTerminal And finally it can be configured and operated using National Instrument s LabView or Lab Windows using the drivers supplied with the unit Application Software operation The SC10 comes with a ...

Страница 9: ...o operate from either the application software or from the front panel controls on the unit The Save Options menu item will save the selected options for future sessions The shutter timing can be changed in any of the following ways using the increment decrement arrows or by entering a time in seconds in the OPEN CLOSED windows When typing a carriage is required to update the SC10 The grey boxes o...

Страница 10: ... and queries must be terminated by a carriage return CR or pressing the ENTER key on the computer The command structure is as follows Keyword argument CR Where keyword defines the function and argument is a numerical value followed by a carriage return CR See listing below The query structure is a follows Keyword CR The keyword defines the function and the question mark Indicates a query The strin...

Страница 11: ...Ex Trigger mode Returns the external trigger mode All commands and queries are in lower case letters If the keyword format or argument are incorrect or out of range the unit will return an error string Software Installation Insert the SC10 CD into the CD ROM drive The program should automatically start installing The install begins by opening install instructions to assist you with the automated i...

Страница 12: ...o be built into a unit by the user e g OEM laser driver cards components mechanics and optics left over parts of units disassembled by the user PCB s housings etc If you wish to return a Thorlabs unit for waste recovery please contact Thorlabs or your nearest dealer for further information Waste treatment on your own responsibility If you do not return an end of life unit to Thorlabs you must hand...

Отзывы: