7
Air Compressor
2.5 Maintenance
Warning
. Do not attempt any maintenance on the compressor until the mains supply has
been isolated and system pressure has been vented. To vent system pressure, operate the
pressure relief valve.
2.5.1 Scheduled Maintenance Frequencies
Caution
. The maintenance tasks detailed in Table 2.1 must be performed within the
periods stated.
Note
. Procedures for performing these tasks are detailed below.
Draining the Air Receiver Unit
During the following procedure, refer to Fig. 2.2 for identification of components.
1) Position the drain tube in a suitable container.
2) Open the drain valve.
3) Close the drain valve when all the water is drained.
Draining the Filter/Regulator Unit
During the following procedure, refer to Fig. 2.2 for identification of components.
1) Position a suitable container below the filter/regulator unit.
2) Locate the drain valve at the bottom of the filter/regulator unit and turn to open the
valve.
3) Close the drain valve when all the water is drained.
Table 2.1 Maintenance Periodicity
Operation
Maintenance period
Drain the Air Receiver
Daily
Check the oil level
Daily
Replenish the oil level
Weekly
Drain the Filter/Regulator
Monthly
Replace oil
Annually
Replace Air Intake filter
Annually
Check pipe fittings and electrical connections
Annually
Check pressure relief activation
Annually