Installation

2.4.3 Setting the Outlet Pressure
Outlet pressure is adjusted by rotating the knob on the Filter/Regulator unit – see 
Fig. 2.2 item 3, clockwise adjustment increases the pressure. The setting can be
locked by depressing the knob.

2.4.4 Using The Compressor
The compressor has a maximum 50% duty cycle, which means that the motor must
not be allowed to run for more than 50% of the time the compressor is in use. The
motor must never be allowed to run continuously otherwise it will overheat and may
become damaged.
Do not ignore air leaks. All air connections must be leak free to prevent the
compressor from overheating.
The motor control box incorporates a thermal overload protector which cuts the power
if the oil temperature is too high – see Fig. 2.1. If the cutout operates, allow the
compressor to cool down (approx 50 mins) before restarting and ensure adequate
airflow around the unit (see Section 2.1 – Siting).

Cautions

The thermal overload protector is set at the factory and should not be adjusted.
You must rectify the cause of the overload before continuing to use the compressor.

Check for:
Drain tap not closed properly
Air leaks on the pneumatic fittings
Excessive workload on compressor, e.g. running too many systems from one
compressor.

Содержание PTA011

Страница 1: ...AIR COMPRESSOR Model Number PTA011 and PTA511...

Страница 2: ...d manufactures building blocks for the photonics industry including equipment for opto mechanics motion control nano positioning alignment optical components laser diodes tunable lasers and vibration...

Страница 3: ...1 Contents INTRODUCTION 2 Safety Precautions 2 INSTALLATION 3 Siting 3 Electrical Connections 3 Preparation 4 Operation 5 Maintenance 7 TROUBLESHOOTING 9 PUMP WIRING DETAILS 11...

Страница 4: ...t you must do Read this manual before using your compressor Ensure the compressor has been correctly installed electrically connected and piped in by a suitably qualified person Ensure the compressor...

Страница 5: ...e To avoid overheating problems do not box in the compressor If sited in a closed room ensure adequate ventilation Do not allow hot air to re circulate If sited outdoors ensure provision of adequate p...

Страница 6: ...taken to avoid contact with the skin Cautions During item 2 use only ISO 32 Fully Synthetic Compressor Oil Do not overfill the reservoir Damage to the oil seals could occur if the level indicated on t...

Страница 7: ...sor When switched on the compressor runs until the maximum working pressure is reached and than automatically switches off As air is used the pressure drops until the minimum operating pressure is rea...

Страница 8: ...not ignore air leaks All air connections must be leak free to prevent the compressor from overheating The motor control box incorporates a thermal overload protector which cuts the power if the oil t...

Страница 9: ...a suitable container 2 Open the drain valve 3 Close the drain valve when all the water is drained Draining the Filter Regulator Unit During the following procedure refer to Fig 2 2 for identification...

Страница 10: ...age to the oil seals could occur if the level indicated on the sight glass is exceeded 5 Refill with oil reservoir until the level is aligned with the mark on the sight glass 6 Refit the oil filler pl...

Страница 11: ...p If leakage persists replace the valve 6 Faulty starter relay Contact the service agency 7 Faulty condenser Replace the condenser or contact the service agency 8 Thermal overload switch active Wait f...

Страница 12: ...pipe connections Compressor does not switch on when pressure is under the minimum level and or does not switch off at maximum pressure 1 Replace pressure switch Compressor gets very hot and or uses a...

Страница 13: ...r 4 Pump Wiring Details Wiring details for the pump unit are shown below Fig 4 1 Pump unit wiring details starter relay overload protector capacitor to pressure switch black black black brown blue yel...

Страница 14: ...12 Intentionally Blank...

Страница 15: ...ser Measurement Instrumentation Laser beam characterization Photodiodes Power and Energy Meters Lab Accessories Technical Support Thorlabs provides a comprehensive after sales service Contact us throu...

Страница 16: ...A 0103T Issue 4 Thorlabs Ltd Saint Thomas Place Ely Cambridgeshire CB7 4EX UK Tel 44 0 1353 654440 Fax 44 0 1353 654444 www thorlabs com Thorlabs Inc 435 Route 206 North Newton NJ07860 USA Tel 1 973 5...

Отзывы: