Page 12
19593-D02
Caution
During item (7) the stage should be homed using the button on the GUI panel.
Do not jog the stage to the zero-position using the jog buttons as this could
establish a false zero datum; there is also a risk that the stage could
overshoot its end stops and become stuck.
Fig. 4.1 APTUser GUI screen
Fig. 4.2 Kinesis GUI screen
7) Move the stage to its home position by pressing “Home” on the GUI, to establish
the zero-datum point. Pay special attention to the Caution note below with
regards to
not
using the jog buttons for this operation.
The stage is now ready for use and/or implementation of parameter changes, as
detailed in the relevant Kinesis/APT manual for the KDC101 controller.