2.2 Functions in the channel menu
MLC8xxx / page 36
The status field shows the actual status:
Normal operation
ON
Laser is on
OFF
Laser is off
Malfunction
Vcc
Internal power supply failed – maintenance required
OTP
Final stage overheated – switch off and wait 10 minutes
ILCK
Interlock is open
ILIM
Limit current reached during operation
In the channel menu you find some abbreviations which are explained as follows
In constant current mode:
I
LS
1 ... I
LS
8
L
aser diode
s
et current of channel 1... 8
I
PA
1 ... I
PA
1
P
hoto diode
a
ctual current of channel 1 ... 8
In constant power mode:
I
PS
1 ... I
PS
8
P
hoto diode
s
et current of channel 1 ... 8
I
LA
1 ... I
LA
8
L
aser diode
a
ctual current of channel 1 ... 8
In both modes
I
LIM
1 ... I
LIM
8
Lim
it current (is the same for chanell 1 ... 8)