14
MAX200 Series Motorized Microscopy Stages
There are various user settings which relate to the position correction modes
discussed above.
1) Click the ‘Settings’ button on the motor GUI, and select the ‘Encoder’ tab.
The lower collection of settings on this panel relate to encoder position
functionality (see Section 3.4.2. for details on the upper collection of settings)..
Fig. 3.5 Motor settings - Encoder tab
There are two encoder positioning correction modes as described previously. The
following parameters are used to adjust the operation of these correction modes.
Consider the following diagram that illustrates the role of each parameter:
Fig. 3.6 Positioning correction mode parameters