17
Chapter 4 Installation
4.2.2 Mounting To An Olympus IX Frame
The stages are mounted to the working surface by M5 screws through the base plate.
1) Remove the four M5 bolts securing the manual XY specimen stage, fitted as
standard to the microscope. Tilt back the lamp support unit as necessary.
2) Remove the manual XY specimin stage.
3) Fit the two shims (provided) between the stage and the microscope. This ensures
that the objective lens and turret cannot clash when the stage is moved.
4) Using the four M5 pozi pan bolts provided, fit the stage to the microscope. To
access the front mounting holes, turn the motor knob and move the carriage until
the holes are accessible in the base plate.
Fig. 4.1 Accessing the stage mounting holes