© 2018 Thorlabs GmbH
32
ITC1xx
·
You don’t get the desired laser operation temperature
Ø
Is the TEC connected properly to the connector?
·
Check all connections and cables.
·
Check the correct
.
Ø
Is the temperature controller turned on?
·
Apply a logical "high" level to pin A13 of ST1 to turn on the laser output. The LED1b
on the controller board must light up.
Ø
Is the temperature sensor connected properly and is the temperature set accordingly?
·
Check the connection and polarity of the temperature sensor - please refer to section
Connecting TEC and Temperature Sensor
.
·
Adjust P4 to the desired temperature (
).
Ø
The TEC current limit might be set to 0?
·
Adjust P3 to the desired TEC current limit (
).
Ø
Is the ITC1xx set correctly for the used temperature sensor?
·
Select the temperature sensor using S3.
·
Set temperature differs from actual temperature (of the laser)
Ø
Is the temperature oscillating?
·
Adjust the PID parameters carefully - please refer to section
.
If an error or malfunction appears that is not listed above, please contact Thorlabs.