![THORLABS EDU-OT2 Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/thorlabs/edu-ot2/edu-ot2_user-manual_1110857059.webp)
Portable Optical Tweezers Kit
Chapter 6: Setup and Adjustment
Page 56
MTN012639-D02
Step 3: Laser Installation
Place the alignment plate into the cage system
right before the 45° mirror, see image to the right.
Insert the laser into the mount without securing it
with the locking screws yet.
Turn the laser on and observe the pattern on the
alignment target. You will see an elliptical shape.
Rotate the collimator in the housing until the long
side of the ellipse is oriented in vertical direction.
Fix the locking screws. Make sure the locking
screws fix the collimator close to the thread, see
image below.
Step 4: Laser Mount Alignment
Turn the tip and tilt screws on the KC1-T mount
until the brightest spot of the laser light
distribution hits the hole in the alignment plate.
Turn off the laser.
Remove the alignment plate.
Step 5: Adjustment Mirror Alignment, Part 1
The second (upper) adjustment mirror
must also be removed from its mount.
To do this unscrew the set screw,
place a piece of adhesive tape on the
back of the mirror and pull it out of the
mount.
Turn on the laser.
Adjust the lower adjustment mirror so
the laser beam exits the upper KCB1
vertically
. For that, using a piece of
paper to view the laser spot helps.
When aligned vertically, the laser beam will exit
close to one side of the mount
aperture
. That is not problematic: when the mirror is installed, the laser will hit the
mirror approximately in the center.
Turn off the laser.
Place the mirror back into the mount and fix the locking screw.