![THORLABS ECL1525-PM Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/thorlabs/ecl1525-pm/ecl1525-pm_operation-manual_1110697033.webp)
6.3
ECL1525 specific commands
ECL1525M-PM/SM / page 29
6.3.5 Case Temperature (CaseTemp)
Purpose: Returns the case temperature.
Synopsis
Register
name
Register
number
Read/
Write
Datatype
Response
Generated
Can be
pending?
Volatile?
Access?
Default
contents
R Unsigned
short
5ms no
Volatile
Temp-
Case
0x88
Returns
Same as the Channel (0x30) command.
Detailed description
11:0
Case Temperature
Bit 11:0
The case temperature is defined as a signed short integer with
units of
°
C*100.
6.3.6 Optical Etalon Ratio (OER)
Purpose: OER returns the ratio between etalon power and the reference optical
power.
Synopsis
Register
name
Register
number
Read/
Write
Datatype
Response
Generated
Can be
pending?
Volatile?
Access?
Default
contents
R Unsigned
short
5ms no
Volatile
OER 0x8A
Returns
Same as the Channel (0x30) command.
Detailed description
13:0
Optical Etalon Ratio
Содержание ECL1525-PM
Страница 1: ...Operation Manual Digitally Controlled Continously Tunable Laser ECL ECL1525M PM SM 2005 ...
Страница 2: ...Version 1 0 Date 16 Feb 05 Copyright 2005 Thorlabs Sweden AB Sweden ...
Страница 6: ...1 1 Safety ECL1525M PM SM page 2 ...
Страница 11: ...2 1 Size ECL1525M PM SM page 7 2 Mechanical Interface 2 1 Size Optical Module ...
Страница 12: ...2 1 Size ECL1525M PM SM page 8 Electronic Control Board ...
Страница 14: ......
Страница 38: ......