Single-Channel Piezo Controller, 75 V to 150 V
Zero
- used to zero the position sensor (strain gauge) when operating in 'Closed Loop
mode' - see Section 5.2.4. Setting The Position Sensor Zero.
Enable
- enables or disables the HV channel's output voltage. With the piezo
enabled, the LED in the button is lit. When disabled, the LED is unlit and the output
voltage on the HV amp channel is set to zero volts.
Digital display
- shows the voltage applied to the piezo, or the position (in microns),
as determined by the feedback signal (if equipped). The display mode is set via the
'Volts' and 'Position buttons.
Volts and Position
- push button controls used to toggle the display between volts and
position modes - see Section 5.2.5.. The corresponding button LED is lit when selected.
Open Loop and Closed Loop
- push button control used to toggle between open
loop and closed loop modes. The corresponding button LED is lit when selected.
Settings display
- shows the following user specified settings:
Jog Step Size
- the distance to move when a jog command is initiated. If in closed
loop mode, the step size is measured in microns; if in open loop mode, the step
size is measured in Volts. The step size can be set either via the Settings panel or
by calling the SetJogStepSize method.
Input Source
- displays the input source associated with the selected channel. The
input source can be set either via the 'Settings' panel or by calling the SetIPSource
method - see Section 5.2.1.
Proportional
- displays the proportional feedback loop constant for the selected
channel (0 to 255).
Integral
- displays the integration feedback loop constant for the selected channel
(0 to 255).
Settings button
- Displays the 'Settings' panel, which allows the tuning parameters
to be entered - see Section 5.2.
Ident
- when this button is pressed, the main digital display on the unit associated with
the GUI panel will flash for a short period.
Active
- lit when the unit is operating normally and no error condition exists.
Error
- lit when one of the following fault condition occurs:
1) One or more power supply voltages are out of range.
2) Closed loop mode is selected but the associated piezo is disconnected.
3) An overtemp condition has occurred - see Section 4.6.
5.2 Settings Panel
When the 'Settings' button on the GUI panel is clicked, the 'Settings' window is
displayed. This panel allows data such as jog step size and input sources to be
entered. Note that all of these parameters have programmable equivalents accessible
through the ActiveX methods and properties on this Control (refer to the
Programming