Rev F Dec 2018
Page 36
Chapter 5 Software Reference
5.2.3 Force Sensing tab
Fig. 5.2 Piezo Settings panel - Force Sensing tab
Force Sense Mode
- Check this box to select Force Sensing Mode.
Force Calibration
- This parameter specifies the calibration factor for the type of force
sensor being used. For example, if set to 1, the GUI digital display shows a detected
force of 0 to 1. The default setting for this parameter is 30, to be compatible with our
FSC102 force sensor, which is specified to read forces up to 30N.
Force Offset
- The force sensor may display an offset when no contact is apparent.
This parameter is used to remove any latent offset, such that the sensor only detects
a force when contact is experienced.
5.2.4 Setting the Position Sensor Zero
The position sensor is a strain gauge fitted to the piezo actuator. Due to limitations in
manufacture, the strain gauge may give a small signal when the actuator is at zero
position with zero volts applied. This 'offset' signal must be removed before the
position attained by a specific applied voltage and the position attained by a
corresponding specified distance can be rationalized. The offset is removed by
clicking the 'Zero' button in the GUI panel or by calling the 'ZeroPosition ' method in
the application software.
Each actuator has a unique offset value. Therefore the offset should be adjusted
whenever an actuator is replaced. Temperature and prolonged use can affect the
performance of the strain gauge and therefore the offset value. It is good practice to
adjust the offset value whenever the unit is powered up.
5.2.5 Choice of Display Mode
It is possible to set the GUI to display either the voltage applied to the actuator or the
position measured by the sensor (in microns). Even if the module is in open-loop
mode, the signal from the position sensor can still be displayed.