Single-Channel Piezo Controller, 75 V to 150 V
3.3 Electrical Installation
3.3.1 Connecting To The Supply
3.3.2 Fuses
Two T 3.15A/250V a.c. antisurge ceramic fuses are located on the back panel, one
for the live feed and one for the neutral.
When replacing fuses:
1) Switch off the power and disconnect the power cord before removing the fuse
cover.
2) Always replace broken fuses with a fuse of the same rating and type.
3.3.3 Powering Down The Unit
Warning: Risk of Electrical Shock
The unit must be connected only to an earthed fused supply of 85V to 264V.
Use only power supply cables supplied by Thorlabs, other cables may not be
rated to the same current. The unit is shipped with appropriate power cables for
use in the UK, Europe and the USA. When shipped to other territories the
appropriate power plug must be fitted by the user. Cable identification is as
follows:
Brown Live
Blue Neutral
Green/Yellow Earth/Ground
Warning
Piezo actuators are driven by high voltages. Voltages up to 150V may be
present at the SMC connector. This is hazardous and can cause serious injury.
Appropriate care should be taken when using this device.
Persons using the device must understand the hazards associated with using
high voltages and the steps necessary to avoid risk of electrical shock.
The piezo controller must be switched OFF before the piezo actuator is plugged
in or unplugged. Failure to switch the controller off may result in damage to
either the controller, the stage or both.
Warning
The controller may cause drive voltage spikes on power down. In applications
requiring the highest sensitivity, the piezo drive voltage should be set to zero,
and the HV cable disconnected before the unit is powered down.