Page 34
Chapter 5 Software Reference
Guide
in the
APTServer helpfile
(accessed via the Windows ‘Start’ menu) for further
details and to Section 1.2.4. for an overview of the APT ActiveX controls).
5.2.1 General tab.
Fig. 5.3 Piezo Settings panel - General tab
Jog Step Size
- the distance to move when a jog command is initiated. If in closed loop
mode, the step size is measured in microns; if in open loop mode, the step size is
measured in Volts.
Feedback Loop
- these parameters determine the response characteristics when
operating in closed loop mode. In general, the default values should not be altered.
Proportional
- sets the proportional feedback loop constant for the selected
channel (0 to 255).
Integral
- sets the integral feedback loop constant for the selected channel
(0 to 255).
Input Source
- determines the input source(s) which controls the output from the HV
amplifier circuit (i.e. the drive to the piezo actuators).
SW Only - the unit responds only to software inputs and the HV amp output is that
set using the SetVoltOutput method (or the GUI panel ‘Output’ control).
BNC (+ SW) - the unit sums the differential signal on the rear panel EXT IN (+)
and EXT IN (-) BNC connectors with the voltage set using the SetVoltOutput
method (or the GUI panel ‘Output’ control).
Pot (+ SW) - the unit sums the voltage set using the front panel ‘Control’
potentiometer, with the voltage set using the SetVoltOutput method (or the GUI
panel ‘Output’ control).
Pot + BNC (+ SW) - the unit sums the voltage set using the front panel ‘Control’
potentiometer, with the differential signal on the rear panel EXT IN (+) and EXT IN
(-) BNC connectors and the voltage set using the SetVoltOutput method (or the
GUI panel ‘Output’ control).