![Thoratec HeartMate II LVAS Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/thoratec/heartmate-ii-lvas/heartmate-ii-lvas_manual_1110579029.webp)
Display Module
When connected to the PBU, the Display Module (
Figure 20
) provides an abbreviat-
ed, small-scale display of system performance. The Display Module reports data
from the System Controller via the PBU. The Display Module displays current pump
mode, speed, and pulse index; as well as pump power and flow data.
When an alarm is active (
Figure 21
), the highest priority alarm message alternates
with the pump power and flow information. Hazard alarms appear in upper case
letters and advisory alarms appear in lower case.
CLINICAL OPERATION AND PATIENT MANAGEMENT
26
Fixed Speed 9600 PI 5.5
LOW FLOW
for < 1 min
Figure 21
Sample Alarm Message on Display Module Screen
Fixed Speed 9600 PI 5.5
Flow 4.5 Power 8.2
Figure 20
Display Module with Sample Display