The Settings Screen contains:
System Status Boxes
Various system parameters are displayed in two System Status boxes.
Active Alarm Messages
The two highest active alarm messages appear as text banners below the oper-
ating mode. If more than two alarms are occurring simultaneously, a “+” will
appear at the end of the second alarm banner, indicating that additional alarms
are active.
Command Buttons
-
The Fixed Speed Adjust button allows users to increase or decrease the fixed
speed in increments of 200 rpm, within a range of 6,000 - 15,000 rpm.
-
The Low Speed Limit button allows users to increase or decrease the low
speed limit in 200 rpm increments between 8,000 and 10,000 rpm. This is
the alarm limit for low speed operation advisory alarm and the speed to
which pump speed will decrease when a suction event is detected.
-
The Pump Stop/Start button allows users to stop or start the pump in a
controlled manner. To stop the pump, press and hold the stop button. A
countdown from 15 will begin. Once the Pump Off alarm message appears,
the pump stop command has been accepted and the stop button can be
released.
-
The Silence Alarm button accompanies any active audio alarms. Pressing
this button silences hazard alarms and Power Cable Disconnected advisory
alarms for two minutes and other advisory alarms for four hours.
CLINICAL OPERATION AND PATIENT MANAGEMENT
21