The Clinical Screen contains:
Parameter Boxes
Four boxes at the top of the screen report measured values of the pump flow,
pump speed, pulsatility index (abbreviated on screen as “Pulse Index”), and
pump power.
Operating Mode and Speed Setpoint
The operating mode and speed setpoint are displayed below the parameter
boxes.
Active Alarm Messages
The two highest active alarm messages appear as text banners below the oper-
ating mode. If more than two alarms are occurring simultaneously, a “+” will
appear at the end of the second alarm banner, indicating that additional alarms
are active.
Command Buttons
(that appear during certain conditions)
-
The Pump Start button appears when the pump is stopped or disconnected
from the System Controller. Pressing this button restarts the pump.
-
The Silence Alarm button accompanies any active audio alarms. Pressing
this button silences hazard alarms and Power Cable Disconnected advisory
alarms for two minutes and other advisory alarms for four hours.
Settings Screen
The Settings Screen (
Figure 15
) allows users to monitor system parameters,
change speed settings, and manually stop or start the pump.
CLINICAL OPERATION AND PATIENT MANAGEMENT
20
Figure 15
Sample Settings Screen