8 - 18
021326/05/10
Tools Required:
- Wrench, wrench width13
- Allen wrench, wrench width 3
Installation is carried out as follows (see also Figure 1
)
1. Remove the precipitation transmitter from the box.
2. Remove the small box from the collecting funnel of the housing
(1)
.
Unpack tipping bucket
(7)
and filter
(12)
and put them aside.
Important:
The tipping bucket is allocated to the precipitation transmitter.
Tipping bucket and precipitation transmitter are marked each by the same tipping bucket
number.
The tipping bucket number is located as small label
- on the small box.
- on the pan
- on the base
3. Remove the bag from the collecting funnel of the housing
(1)
,
remove the combination filter/cram
(14/15)
and put it aside.
4. Put the precipitation transmitter on a suitable base, and fix it by means of the screws
(4)
inside the base
(3)
.
5. Loosen two screws
(2)
at the housing
(1)
and lift the housing carefully up.
6. Please check, if the level- indicator
(5)
on the ground plate indicates a horizontal
mounting, if necessary, correct it by the screws
(4)
inside the base
(3)
.
7. Push the collector
(13)
upward by loosening the knurled screw
(11)
and fix it.
8. Then insert the tipping bucket
(7)
carefully into the bearing cup of the pan
(6)
.
Important:
- The magnet
(10)
of the tipping bucket must indicate towards the pc-board.
- Please check once again if the tipping bucket, and the precipitation transmitter
have the right allocation (synchronize numbers).
9. Afterwards, test the tipping bucket manually on trouble-free tipping.
10. Insert the filter
(12)
, which was put aside, into the collector
(13)
.
11. Put the collector back into lower position, and fix it.
12. Put the housing carefully back on the instrument, and fix it by means of the two
instrument screws.
13. Insert the combination filter/cram
(14/15)
, which was put aside, into the passage of the
collecting funnel.