• Disconnect the therapy wrap from the unit completely. Re-connect the umbilical to the
unit and re-connect the wrap to the umbilical. Visually check fluid flow. Let the system
run for 1 minute or so to cool the wrap which determines fluid flow.
22
23
Re-connect the wrap to the unit. Check the fluid flow
and verify all air bubbles have been purged.
Disconnect therapy wrap from unit completely.
Re-connect the umbilical to the unit.