![ThermoTek VascuTherm 4 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/thermotek/vascutherm-4/vascutherm-4_user-manual_1106306058.webp)
Internal fault within the therapy
unit.
Check if the power supply LED
is green.
Turn the unit off by
disconnecting the power
supply. Wait 10 minutes and
reconnect the power supply.
Try an alternate AC outlet.
Contact ThermoTek Customer
Service if problem persists.
Problem/Alarm Screen
Cause
Suggested Actions
The touch screen is not
functioning.
Physical damage to unit. Inspect the unit for physical damage.
If the unit shows any cracks or dents,
the unit should not be used. Contact
the clinic or DME for assistance.
Unit not connected to
AC power.
Make sure the unit is connected to
the power supply and that the power
supply is connected to an appropriate
AC outlet.
Internal fault within the
therapy unit.
Check if the power supply LED is green.
Turn the unit off by disconnecting the
power supply. Wait 10 minutes and re-
connect the power supply.
Try an alternate AC outlet.
Contact ThermoTek customer service if
problem persists.
Problem/Alarm Screen
Cause
Suggested Actions
My unit gives me a TIMER NOT
SET alarm.
The timer on your unit
has not been set to run
cool or heat therapy.
Contact your provider so they can
program your unit.
54
Содержание VascuTherm 4
Страница 1: ...VascuTherm 4 User Manual...
Страница 2: ......
Страница 12: ...Observe all warning and caution labels Never remove the caution warning labels 8...
Страница 76: ...This page intentionally left blank...
Страница 77: ...This page intentionally left blank...
Страница 78: ...This page intentionally left blank...
Страница 79: ...This page intentionally left blank...