8
Set up a program
v
To set up
The program setting button is located on the upper right of the main
screen, and the method for setting the program according to the usage
environment is as follows.
§
Place the metric reference article at 1.0m or 1.5m from the instrument.
§
Click the program setting button and set the frequency on the setting
side.
※ If you do not press the [Apply] button, changes will not be applie
to the program.
§
Check the degree of the measurement reference through the screen.
§
Measure the degree of the measurement reference using a non-
contact thermometer.
§
Repeat the above three sections 3 times to 5 times.
§
Check the deviation value of the degree measured by the thermal
camera compared to the degree measured by the non-contact
thermometer.
§
Click the program setting button and set the degree deviation value
as input in the degree correction section on the setting menu.
※ If you do not press the [Apply] button, changes will not be applied
to the program.