General guidelines for
decontaminating Pipettes when
Working with Different Liquids
g
q
Liquid
Handling, Special features
Decontamination
Aqueous
Pipettes are calibrated with
Open pipette, rinse contaminated parts
solutions and
distilled water. Results are
well with distilled water, and allow to dry
buffers
extremely accurate.
at maximum 60
°
C in dryer compartment.
Lubricate piston if neccessary using the
grease that comes with the pipette.
Inorganic acids
It is advisable to occasionally rinse
The plastics used in Thermo Labsystems
the lower part of the pipette with
pipettes are acid resistant. However,
distilled water if high-concentration
aerosols from the acids can enter the
acids are pipetted frequently. Using
lower part of the pipette and affect the
Filtertips is also recommended.
performance of the pipette. Clean as described
above in “Aqueous solutions and buffers”.
Alkalis
It is advisable to occasionally rinse the
The plastics used in Thermo Labsystems
lower part of the pipette with distilled
pipettes are alkali-resistant. However,
water if high-concentration alkalis are
aerosols from the alkalis can enter the lower
pipetted frequently. Using Filtertips is
part of the pipette and affect the perform-
also recommended.
ance of the pipette. Clean as described
above in “Aqueous solutions and buffers”.
Potentially
To avoid contamination, Filtertips should
Use 70% ethyl alcohol to disinfect.
infectious liquids
be used. Alternatively, positive
The tip cone and tip ejector can be
displacement systems can be used.
left in an ethyl alcohol bath overnight.
Autoclave the Finnpipette Digital in one
piece at 121
°
C for 20 min. Viruses and
spores can be inactivated by wiping the
the pipette with a tissue moistened with
5% sodium hypochlorite. Moisten a clean
tissue with distilled water and wipe the
surface well.
Cell cultures
To guarantee sterility, Thermo Labsystems Proceed as described above with
Filtertips should be used.
“potentially infectious liquids”.
Organic solvents
1.Density is different than that of water.
This evaporation process is normally
Therefore, it is necessary to adjust
sufficient for liquids with a high vapour
the pipette.
pressure. Alternatively, immerse the
2.Pipetting should be carried out rapidly,
contaminated parts in detergent. Rinse
due to the high vapor pressure and
well with distilled water and dry as
changes in the wetting pressure.
described above.
3.After pipetting is finished, open
the pipette and allow the liquid
to evaporate.
Radioactive
To avoid contamination, Filtertips should
Open pipette and place contaminated
solutions
be used. An alternative would be to use
parts in complex solutions or special
positive displacement systems.
cleaning solutions. Rinse well with distilled
water and dry as described above.
Proteins
To avoid contamination, Filtertips should
Open pipette, rinse pipette with
be used. An alternative would be to use
detergent. Rinse well with distilled
positive displacement systems.
water and dry as described above.
Lightly lubricate piston.
Nucleic acids
To avoid contamination, Filtertips or
If you have an autoclavable pipette,
positive displacement systems should
autoclave it according to the manufactorer's
be used.
instructions. Otherwise open pipette, wipe
with 90% ethanol, followed by 2M sodium
acetate and again 90% ethanol.