19
Manual multichannel pipettes
-
The pipette is held in the calibration room for at least 2 hours before calibration
to reach equilibrium with the test room conditions.
-
The pipette is checked at the maximum volume (nominal volume) and at the
minimum volume or 10% of the maximum volume, whichever is higher.
For example, Finnpipette 0.5 - 10
µ
l is tested at 10
µ
l and 1
µ
l.
-
Both volumes are tested with the two end channels.
-
A series of ten pipettings is performed with both volumes.
-
Calculate the accuracy and precision using the formulas below. If the calculated
results are within the limits given in the Instructions-for-use booklet, the pipette
calibration is correct. If not, the pipette has to be adjusted with the lower
volume (see below) and both end channels have to be checked again.
Finnpipette BioControl single channel module
-
The pipette is held in the calibration room for at least 2 hours before calibration
to reach equilibrium with the test room conditions.
-
Use the Finnpipette BioControl calibration software that comes with
the pipette.
-
The pipette is checked at the maximum volume (nominal volume) and at the
lower calibration volume, which can be found from the Instructions-for-use
booklet or from the Finnpipette BioControl calibration software.
-
A series of ten pipettings is performed with both volumes.
-
Calculate the accuracy and precision of both series using the formulas below
and compare to the limits given in the “Checking the calibration” chapter in the
Instructions-for-use booklet. If the calculated results are within the selected
limits, the pipette calibration is correct. If not, adjust the pipette as
described below.
Finnpipette BioControl multichannel module
-
The pipette is held in the calibration room for at least 2 hours before calibration
to reach equilibrium with the test room conditions.
-
The pipette is checked at the maximum volume (nominal volume) and at the
lower calibration volume, which can be found from the Instructions-for-use
booklet or from the Finnpipette BioControl calibration software.
-
Both volumes are tested with the two end channels.
-
A series of ten pipettings is performed with both volumes.
-
Calculate the accuracy and precision of both series using the formulas below
and compare to the limits given in the “Checking the calibration” chapter in the
Instructions-for-use booklet. If the calculated results are within the selected
limits, the pipette calibration is correct. If not, adjust the pipette as
described below.
Procedures to adjust the pipette:
Manual single channel pipettes
-
The adjustment is done at the lower volume.
-
Place the service tool that comes with the pipette into the openings of the