9 Section 2 | Overview
Flask size
Size
Liquid volume
Max Speed(rpm)
50ml
≤
50ml
550
100ml
≤
100ml
550
150ml
≤
150ml
550
550
200ml
≤
200ml
250ml
≤
250ml
500
500ml
≤
500ml
400
1000ml
≤
1000ml
280
Max Flask
Capacity
49
25
25
18
18
10
7
≤
500ml
480
7
2000ml
≤
2000ml
230
4
≤
1000ml
430
4
3000ml
≤
3000ml
200
4
≤
1500ml
430
4
5000ml
≤
5000ml
150
≤
2500ml
280
2
2
Note:
1) The flask clamps (
≤
2L) can be placed anywhere on the universal (aluminum)
platform, even if the load is not balanced.
▲
2) Please refer to the "Liquid Volume" table in the specifications section for the
maximum load.
▲
Table 2-4.
Maximum Shaking Speed and Flask Capacity on 18 inch (aluminum)
Platform
Note:
1) The flask clamps (<2L) can be placed anywhere on the 18inch (aluminum)
platform, even if the load is not balanced.
▲
2) Please refer to the "Liquid Volume" table in the specifications section for the
maximum load.
▲
Section 3 Operation
3.1 Operations on the Control Panel
The unit has been designed with a “soft
start/slow down” providing a
smooth acceleration/ deceleration
during operation. This feature will
decrease splashing of solutions from
vessels during operation.
Warning:
The microprocessor control
system may take 1 minute to
synchronize the actual RPM of the
universal (aluminum) platform to the
desired RPM setting. Make sure that all
the flasks and test tubes are firmly
fixed on the universal platform and the
flask clamps are fasten securely
with the screws.
①
②
⑩
③
④
⑤
⑥
⑦
⑧
⑨
1) Display LED: Shows time or speed
2) RPM indicator light: LED shows
current speed
3) TIME indicator light: LED shows
current time
4) WARN indicator light: Over-speed
alarm
5) RUN indicator light: The instrument is
running
6) RPM/TIME Key: While instrument is in
operation, press to change time and
speed display; when setting
parameter, press to change minutes,
hour or speed.
7) Up Arrow Key: Increases value. Hold
for 3 seconds to increase rapidly
8) Down Arrow Key: Decreases value.
Hold for 3 seconds to decrease
rapidly.
9) Key: The power is On/Off.
10) Run/Stop Key: The instrument is
running or stopped.
MaxQ 2000 CO2 Plus 10