manualshive.com logo in svg
background image

 
 

Doc No.  

November 2018 

 

 

III.

 

Troubleshooting Tips 

Flow restriction or blockage from detector cell or backpressure coils should be eliminated 
first to avoid irreversible damage to the suppressor.  

 

Detector Cell Backflush: 

Bypass the suppressor and connect the column 

outlet

 tubing to the detector 

cell 

outlet

 port and direct the detector cell 

inlet 

tubing to waste. 

 

 

 

Turn on the pump and flow eluent at the application flow rate for 10 minutes. 
Turn off the pump.  Re-measure the detector cell backpressure by following 
the steps in 

Section I

 above. 

 

 

Note:

  Repeat  the  backflush  procedure  if  the  backpressure  remains  high.  If  the 

backpressure  of  the  detector  cell  doesn’t  come  below 

70  psi  after  multiple  attempts, 

replace the detector cell. 
 

             

Backpressure Coil Adjustment: 

If the pressure is greater than 40 psi for carbonate eluent or other eluents, 
trim the backpressure coil and repeat the backpressure coil test in 

Section I

.  

 

If it is less than 30 psi for carbonate eluent or less than 10 psi other eluents, 
add more tubing and repeat the backpressure coil test in 

Section I

 

 

Note:

 Backpressure coils help to prevent gases generated during AutoSuppression from 

out-gassing by compressing the bubbles. The bubbles can be occasionally trapped in the 
detector cell causing noise. For example, carbonate eluent is suppressed to carbonic acid, 
which is CO

2

 gas in equilibrium with DI water, and CO

2

 gas can come out of solution if 

adequate pressure is not applied. Therefore, we highly recommend the addition of 30-40 
psi for carbonate eluent or 10-40 psi for other eluents. 

 

 
Suppressor Backflush with DI Water: 

Turn off the pump and disconnect the column 

outlet

 and cell 

inlet

 tubing from 

the suppressor eluent ports (if not already disconnected).   

 

Connect the DI water tubing from the pump 

outlet 

directly to the suppressor 

eluent

 

out

 port and direct the 

eluent in

 tubing to waste. (Effectively plumbing 

the eluent ports backwards to reverse flow through the eluent chamber).   

 

Turn on the pump and flow DI water at 1.0 mL/min for 4 mm suppressor and 
0.25 mL/min for 2 mm suppressor for 10 minutes. Turn off the pump. Re-
measure the suppressor backpressure by following the steps in 

Section II

.  

 

 

Note:

  Repeat  the  backflush  procedure  if  the  backpressure  remains  high.  If  the 

backpressure  of  the  suppressor  doesn’t  come  below  200  psi  after  multiple  attempts, 

replace the suppressor or contact your Service Engineer or Technical Support team.

 

 

Отзывы: