November 2018 

 

 

Allow  the  suppressor  to 

rest  undisturbed

  for  20  minutes  to  fully  hydrate  the 

resin, screens, and membranes. 

 

Turn on the pump at the application flow rate for 10 minutes; record the system 
pressure 4 (

P

4

) and turn off the pump.  

 

P

4

, _______ psi 

 

Calculate the backpressure across the suppressor (

BP

3

):  

             

BP

3

 = 

P

4

 

 

P

1

 = ________ psi                                                     

Pass:

 

≤ 

200 psi 

 
If the backpressure exceeds this specification, see 

Suppressor Backflush with DI 

Water

 section 

 

 

                                                

 

Step 5:

 Connect the suppressor 

Eluent outlet

 tubing to the detector cell

 inlet.

 

Then  connect  the  detector  cell

  outlet

  tubing  with  backpressure  coil(s)

 

to  the 

Regen Inlet

 of the suppressor, as shown in 

Figure 2

, for the Auto Suppression 

Recycle Mode Configuration. 

For External Water Mode Configuration, refer to 

the suppressor manual for detailed installation guide. 

 

The suppressor is now ready to be installed and configured (see 

Configuration 

of the New Suppressor

 below). Normal operation using the application eluent 

strength, flow  rate,  and recommended  current/voltage  power  settings  can  be 
applied. 

 

 

 
 

 

Figure 2:

 

AutoSuppression Recycle Mode Configuration

 